The RMS-based multilead delineation algorithm performed equivalently to the best single-lead delineation for the 2-lead QT database, operating within a fraction of a sample duration of the CSE committee tolerances.
Optimized RMS-based multilead DWT algorithms enable accurate, real-time ECG wave delineation on low-power wearable sensor nodes.
This work is devoted to the evaluation of multilead digital wavelet transform (DWT)-based electrocardiogram (ECG) wave delineation algorithms, which were optimized and ported to a commercial wearable sensor platform. More specifically, we investigate the use of root-mean squared (RMS)-based multilead followed by a single-lead online delineation algorithm, which is based on a state-of-the-art offline single-lead delineator. The algorithmic transformations and software optimizations necessary to enable embedded ECG delineation notwithstanding the limited processing and storage resources of the target platform are described, and the performance of the resulting implementations are analyzed in terms of delineation accuracy, execution time, and memory usage. Interestingly, RMS-based multilead delineation is shown to perform equivalently to the best single-lead delineation for the 2-lead QT database (QTDB), within a fraction of a sample duration of the Common Standards for Electrocardiography (CSE) committee tolerances. Finally, a comprehensive evaluation of the energy consumption entailed by the considered algorithms is proposed, which allows very relevant insights into the dominant energy-draining functionalities and which suggests suitable design guidelines for long-lasting wearable ECG monitoring systems.
Rincón et al. (Mon,) conducted a other in Electrocardiogram (ECG) monitoring (n=105). RMS-based multilead wavelet-based ECG delineation algorithm vs. Single-lead delineation algorithm was evaluated on Delineation accuracy (mean error and standard deviation of characteristic wave boundaries). The RMS-based multilead delineation algorithm performed equivalently to the best single-lead delineation for the 2-lead QT database, operating within a fraction of a sample duration of the CSE committee tolerances.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: