March 1, 1989Pacing and Clinical Electrophysiology26 citations

Treadmill Assessment of an Activity‐Modulated Pacemaker: The Importance of Individual Programming

View Full Paper
HMHugh McAlisterJSJudith E. SobermanPKPeter T. Klementowicz

Key Points

Key points are not available for this paper at this time.

Abstract

Maximum benefit from a rate-modulated pacemaker requires individualized programming of rate response settings. We tested an externally strapped activity-sensing pacemaker (Activitrax-Medtronic 8400) in eight healthy volunteers, to assess the pacing responses of the different rate response and activity threshold settings. Five males and three females, aged 20 to 70 years (mean 40), performed a total of 67 treadmill exercise tests, using a specific protocol designed to assess the activity-sensing unit. The external unit was compared to implanted units in four patients, to validate its accuracy. A reproducible sinus response to the treadmill protocol was observed, against which pacing responses were compared. The activity threshold determines the degree of activity required to elicit a pacing rate response, whereas the rate response setting determines the rate attained. Rates of 140 bpm were rarely achieved, despite vigorous exercise. The sensor responds rapidly to activity, not to physiologic demand; to increase in speed, not grade. Four patients performed repeated limited treadmill tests to determine their optimum program setting, with symptomatic status and the healthy volunteer sinus response as guides. These results, and those from the external Activitrax unit, suggest that LOW 6 and MEDIUM 6-10 settings will prove optimum for most patients.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

McAlister et al. (1989) studied this question.

synapsesocial.com/papers/6a70c9c5660549caf2c56a3dhttps://doi.org/10.1111/j.1540-8159.1989.tb02686.x
Ask AI
Helpful
Bookmark
Share
View Full Paper

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Limitations of Rate Response of an Activity‐Sensing Rate‐Responsive Pacemaker to Different Forms of Activity1988 · 81 citations
  2. 2Clinical Experience with an Activity Sensing Pacemaker1986 · 47 citations
  3. 3Clinical Experience with an Activity Sensing DDDR Pacemaker Using an Accelerometer Sensor1992 · 41 citations
  4. 4Activity‐Sensing, Rate‐Responsive Pacing: Improvement in Myocardial Performance with Exercise1985 · 147 citations
  5. 5Assessment of the Work Capacity and Relationship Between Rate Response and Exercise Tolerance Associated With Activity‐Sensing Rate‐Responsive Ventricular Pacing1987 · 35 citations