We use a combined electromagnetic-electrophysiological modeling framework to predict cardiac stimulation (CS) thresholds in individualized porcine body models and compare those simulations to thresholds measured in eight pigs using strong dB/dt pulses. For all pigs, the simulated and measured thresholds agree within 30%, and no significant differences between simulations and measurements were detected (p<0.05, paired t-test). The threshold model uncertainty was found to be ~25% in a sensitivity analysis of the relevant model parameters. A well-validated model may help inform appropriate safety limits for MRI gradients to protect patients from CS without overly restricting gradient performance.
Klein et al. (Wed,) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: