Oxygen Tension Under a Hard, Gas‐Permeable Contact Lens

View Full Paper
IFIrving FattDLDoris Lin

Key Points

  • The study aims to calculate the oxygen tension under hard, gas-permeable contact lenses.
  • Calculated oxygen tension using gas permeability data from engineering literature.
  • Compared engineering data with optometric claims.
  • Conducted new measurements of gas permeability.
  • Gas permeability of the lenses was determined to be 10 times smaller than claimed.
  • Calculations indicated oxygen tension is at the lower end of acceptable values.

Abstract

ABSTRACT Oxygen tension under a hard, gas‐permeable contact lens is calculated. The gas permeability of the lens is taken from the engineering literature and is shown to be 10 times smaller than claimed in the optometric literature. The engineering data are supported by new measurements of gas permeability. Calculations using the new data show that the oxygen tension under the lens is at the lower end of the range of acceptable values.