A plug-based bail-out strategy combining an additional TEER device and an Amplatzer Vascular Plug II successfully reduced recurrent mitral regurgitation to mild following a leaflet tear.
Does a plug-based bail-out strategy reduce mitral regurgitation in a patient with a leaflet tear after mitral TEER?
A plug-based strategy combining an additional TEER device and a vascular plug can be an effective transcatheter bail-out option for leaflet tears following mitral TEER in patients at high surgical risk.
Absolute Event Rate: 0% vs 0%
Abstract Background Mitral transcatheter edge-to-edge repair (TEER) is a safe and widespread treatment approach for severe mitral regurgitation (MR). However, rare but serious device-related complications, including single leaflet device attachment, perforation, and leaflet tear, may occur, often requiring surgical correction. Case summary An 82-year-old man with congestive heart failure and ischemic cardiomyopathy underwent mitral TEER using MitraClip XTW clip for severe functional MR. Two days after the procedure, MR recurrence due to a leaflet tear was observed. Given the patient’s advanced age and reduced left ventricular function, our cardiac team decided to perform a transcatheter bail-out procedure. First, the tear was converted to a hole shape by implanting a PASCAL Ace device (Edwards Lifesciences) lateral to the XTW, as this device has a lower risk of leaflet damage compared to the MitraClip device. Then, Amplatzer Vascular Plug II (AVP II) was deployed into the hole, with its mid-lobe positioned in contact with the devices and leaflets. After releasing the AVP II, the MR jet reduced to mild. Conclusion An optimal bail-out strategy for leaflet tears has not yet been established. When it is difficult to sufficiently cover the leaflet tear using only an additional TEER device, the use of a plug with additional TEER devices can be considered a bail-out strategy.
Nishiura et al. (Fri,) reported a other. A plug-based bail-out strategy combining an additional TEER device and an Amplatzer Vascular Plug II successfully reduced recurrent mitral regurgitation to mild following a leaflet tear.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: