Ablation of atrial fibrillation with a cryoballoon compared to a radiofrequency PVAC resulted in similar rates of complete freedom from AF at 12 months (46% vs. 34%, P=0.2).
RCT (n=110)
randomized
paroxysmal or persistent atrial fibrillation (n=110)
Cryoballoon ablation vs Circular multipolar duty-cycled radiofrequency-based pulmonary vein ablation catheter (PVAC)
Complete freedom from AF, without antiarrhythmic drugs, after one single ablation procedure at 12 months, p=0.2
Absolute Event Rate: 46% vs 34%
p-value: p=0.2
AIMS: The urge to facilitate the atrial fibrillation (AF) ablation procedure has led to the development of new ablation catheters specifically designed as 'one-shot tools' for pulmonary vein isolation (PVI). The purpose of this study was to compare the efficacy, safety, and procedure times for two such catheters using different energy sources. METHODS AND RESULTS: One hundred and ten patients, referred for ablation of paroxysmal or persistent AF, were randomized to treatment with either the cryoballoon or the circular multipolar duty-cycled radiofrequency-based pulmonary vein ablation catheter (PVAC). Complete PVI was achieved in 98 vs. 93% patients in the cryoballoon and PVAC group, respectively, with complication rates of 8 vs. 2% (P = 0.2). Complete freedom from AF, without antiarrhythmic drugs, after one single ablation procedure was seen in 46% in the cryoballoon vs. 34% after 12 months (P = 0.2). Procedure times were comparable, but fluoroscopy time was shorter for the cryoballoon (32 ± 16 min) than for the PVAC procedures (47 ± 17 min) (P < 0.001). A significant improvement of quality of life (QoL) and arrhythmia-related symptoms was seen in both groups after ablation. CONCLUSION: Both catheters proved comparably effective and safe in achieving acute PVI, apart from the shorter fluoroscopy times achieved with the cryoballoon. At follow-up, there was no statistically significant difference between the groups regarding freedom from AF and clinical success. The QoL increased to the same levels as for the general Swedish population in both groups.
Building similarity graph...
Analyzing shared references across papers
Loading...
Helena Malmborg
Uppsala University
Stefan Lönnerholm
Uppsala University Hospital
Per Blomström
Electrophysiology
EP Europace
Uppsala University
Building similarity graph...
Analyzing shared references across papers
Loading...
Malmborg et al. (Wed,) conducted a rct in paroxysmal or persistent atrial fibrillation (n=110). Cryoballoon ablation vs. Circular multipolar duty-cycled radiofrequency-based pulmonary vein ablation catheter (PVAC) was evaluated on Complete freedom from AF, without antiarrhythmic drugs, after one single ablation procedure at 12 months (p=0.2). Ablation of atrial fibrillation with a cryoballoon compared to a radiofrequency PVAC resulted in similar rates of complete freedom from AF at 12 months (46% vs. 34%, P=0.2).
synapsesocial.com/papers/6a1109136da82ae745f336ca — DOI: https://doi.org/10.1093/europace/eut104
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: