The 32-channel SGBT transceiver array enables highly accelerated, high spatial resolution cardiac MRI at 7.0T.
Purpose Design, implementation, evaluation, and application of a 32‐channel Self‐Grounded Bow‐Tie (SGBT) transceiver array for cardiac MR (CMR) at 7.0T. Methods The array consists of 32 compact SGBT building blocks. Transmission field ( ) shimming and radiofrequency safety assessment were performed with numerical simulations and benchmarked against phantom experiments. In vivo efficiency mapping was conducted with actual flip angle imaging. The array’s applicability for accelerated high spatial resolution 2D FLASH CINE imaging of the heart was examined in a volunteer study (n = 7). Results shimming provided a uniform field distribution suitable for female and male subjects. Phantom studies demonstrated an excellent agreement between simulated and measured efficiency maps (7% mean difference). The SGBT array afforded a spatial resolution of (0.8 × 0.8 × 2.5) mm 3 for 2D CINE FLASH which is by a factor of 12 superior to standardized cardiovascular MR (CMR) protocols. The density of the SGBT array supports 1D acceleration of up to R = 4 (mean signal‐to‐noise ratio (whole heart) ≥ 16.7, mean contrast‐to‐noise ratio ≥ 13.5) without impairing image quality significantly. Conclusion The compact SGBT building block facilitates a modular high‐density array that supports accelerated and high spatial resolution CMR at 7.0T. The array provides a technological basis for future clinical assessment of parallel transmission techniques.
Eigentler et al. (Thu,) studied this question.