February 1, 2001Blood Pressure Monitoring22 citations

Effect of altering vasoactivity on the measurement of finger blood pressure

View Full Paper
KJK. JagomägiRRRein RaamatJTJaak Talts

Key Result

Oscillometric blood pressure measurement significantly overestimated finger mean blood pressure compared to the Finapres method during peripheral vasoconstriction (mean difference 10.6 mmHg, P<0.01).

Study Design

Type

Cross-Sectional (n=8)

Structured PICO

Does peripheral vasoconstriction affect the agreement between oscillometric and volume clamp methods for measuring finger blood pressure in healthy volunteers?

P
Population
8 healthy volunteers undergoing simultaneous finger blood pressure monitoring using oscillometric and Finapres methods.
E
Exposure
Modified oscillometric method for non-invasive beat-to-beat finger mean arterial pressure monitoring
C
Comparator
Volume clamp method (Finapres) for non-invasive beat-to-beat finger mean arterial pressure monitoring
O
Outcome
Difference in mean arterial pressure (MAPo-MAPf) during episodes with and without peripheral vasoconstrictionsurrogate

Peripheral vasoconstriction causes significant disagreement between oscillometric and volume clamp methods for finger blood pressure measurement, highlighting the need to avoid vasoconstriction during such measurements.

Main Result

Mean Difference: 10.6

p-value: p=<0.01

Abstract

BACKGROUND: Finapres monitors and oscillometric sphygmomanometers are widely used in blood pressure measurements on the fingers. However, the reliability of finger blood pressure measurement devices still remains a matter of debate. DESIGN: The volume clamp and modified oscillometric methods for non-invasive beat-to-beat finger mean arterial pressure monitoring are compared during intensive spontaneous changes in vascular tone. The degree of vasoconstriction is established by recording the thumb pulp skin blood flow with a laser Doppler instrument. The oscillometric mean arterial pressure (MAPo) and the Finapres mean arterial pressure (MAPf) are simultaneously recorded from adjacent fingers in eight healthy volunteers in a sitting position at room temperature 22-23 degrees C. RESULTS: The changes in blood pressure were similarly tracked by the two blood pressure monitors, except the episodes with peripheral vasoconstriction. The difference (MAPo-MAPf) for all simultaneously recorded mean blood pressure values in episodes without vasoconstriction was (mean+/-SD) 0.7+/-1.8mmHg (P=0.33) and in episodes with vasoconstriction 10.6+/-5.6mmHg (P<0.01). A disagreement between the devices during vasoconstriction is assumed to be caused mainly by the tendency of the oscillometric method to overestimate the finger mean blood pressure, and by the tendency of the volume clamp method (Finapres) to underestimate the finger mean blood pressure in condition of peripheral vasoconstriction. CONCLUSION: For both types of finger monitors (oscillometric blood pressure devices as well as Finapres or Portapres) it is recommended that intensive vasoconstriction in the subject be avoided during measurements. The presence of acute vasoconstrictions can be determined by simultaneous recording of finger skin blood flow.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Jagomägi et al. (2001) conducted a cross-sectional in Healthy volunteers (n=8). Oscillometric blood pressure measurement vs. Finapres blood pressure measurement was evaluated on Difference in mean arterial pressure (MAPo-MAPf) during peripheral vasoconstriction (MD 10.6 mmHg, p=<0.01). Oscillometric blood pressure measurement significantly overestimated finger mean blood pressure compared to the Finapres method during peripheral vasoconstriction (mean difference 10.6 mmHg, P<0.01).

synapsesocial.com/papers/6a2222f73081c2f8f8e24771https://doi.org/10.1097/00126097-200102000-00006