The Tiba Medical Ambulo 2400 monitor passed ISO and BHS criteria, demonstrating a mean difference from manual measurement of -1.8±6.5 mmHg for SBP and 0.7±7.3 mmHg for DBP.
Observational (n=85)
Does the Tiba Medical Ambulo 2400 ambulatory blood pressure monitor accurately measure blood pressure compared to manual measurement in children, adolescents, and adults?
The Tiba Medical Ambulo 2400 ambulatory blood pressure monitor is accurate and meets ISO and BHS validation criteria for use in children, adolescents, and adults.
Mean Difference: -1.8
OBJECTIVE: We evaluated the Tiba Medical Ambulo 2400 ambulatory blood pressure monitor accuracy by both the International Standards Organization (ISO) Standard and the British Hypertension Society (BHS) protocol. METHODS: We tested children, adolescents, and adults. The requirements for both ISO and BHS were completed for 85 individuals each. RESULTS: The Ambulo 2400 passed both ISO and BHS criteria. For ISO method 1, the mean±SD for the difference between device and manual systolic blood pressure (SBP) was -1.8±6.5 mmHg; for diastolic blood pressure (DBP) the values were 0.7±7.3 mmHg. For the method 2 analyses, the differences were -1.8±4.7 and 0.7±6.3 mmHg for SBP and DBP, respectively. The percentage of differences within 5, 10, and 15 mmHg were 60, 86, and 98% for SBP, and 60, 85, and 96% for DBP. CONCLUSION: The Ambulo 2400 passed all phases of the ISO testing and achieved an A grade for both SBP and DBP per BHS analyses. There was no decrease in accuracy in either high or low BP ranges. The Ambulo 2400 integrates activity monitoring for assigning actual awake and asleep times. There are software features that are particularly beneficial to monitor children and adolescents.
Bruce S. Alpert (Thu,) conducted a observational in Blood pressure monitoring (n=85). Tiba Medical Ambulo 2400 ambulatory blood pressure monitor vs. Manual blood pressure measurement was evaluated on Difference between device and manual systolic and diastolic blood pressure (MD -1.8 mmHg for SBP). The Tiba Medical Ambulo 2400 monitor passed ISO and BHS criteria, demonstrating a mean difference from manual measurement of -1.8±6.5 mmHg for SBP and 0.7±7.3 mmHg for DBP.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: