Abstract

The ionization behavior of stable radicals (1–8) was studied under electrospray ionization (ESI) and atmospheric pressure ionization (APCI) conditions. With APCI, using methanol + water as the solvent, most radicals were protonated to give the radical cation M + 1+•, whereas TEMPO (5) was detected as the M+ ion and galvinoxyl (1) gave the M + 2+ ion. Using benzene as the solvent, all radicals (1–8) were efficiently oxidized to give M+ as the base peak. Under ESI conditions using methanol + water with 1% acetic acid, stable radicals (5–7) gave M + 1+•, (2–4)M+, whereas (1) and (8) showed the M + 2+ ion as the base peak. Using the ESI source as a controlled-current electrolytic cell, radicals (2–8) were efficiently oxidized to give M+ ions. Radical (1) was detected as the M + 2+ ion. Detection limits for the radicals (1–8) were between 3 × 10−9 mol L−1 and 5 × 10−7 mol L−1 using APCI or ESI in the full-scan mode. In most cases detection limits were lower using the tandem mass spectrometry (MS/MS) full-scan mode.