The A a level of agreement which would receive a 'B+' grade from the recent British Hypertension Society guidelines. The limits of agreement (2 standard deviations about the mean difference) for systolic BP between observers and the TM-2420 tended to be greater for the standing position (-20 to 15 mm Hg) compared to supine (-14 to 12 mm Hg) and seated (-13 to 8 mm Hg) positions. Limits of agreement between the observers and device were not dependent upon age, heart rate, arm size, or blood pressure level. Twenty-four-hour blood pressure monitoring in two of the four centers demonstrated an error code rate of 3.4%, excluding 'retries' that are one of the device's features. These data demonstrate an acceptable level of accuracy and performance of the sixth generation of the TM-2420 for use in clinical practice and research.(ABSTRACT TRUNCATED AT 250 WORDS)