Transvenous laser-assisted lead extraction achieved a 97.7% clinical success rate and 96.5% complete removal rate, with a 1.4% rate of procedure-related major adverse events.
Cohort (n=1,449)
Yes
What are the safety, efficacy, and risk factors associated with transvenous laser-assisted lead extraction in patients with cardiovascular implantable electronic devices?
Transvenous laser-assisted lead extraction is highly successful with a low procedural complication rate, though mortality is substantially increased in patients with pocket infections, device-related endocarditis, diabetes, or renal insufficiency.
OBJECTIVES: This study sought to examine the safety and efficacy of laser-assisted lead extraction and the indications, outcomes, and risk factors in a large series of consecutive patients. BACKGROUND: The need for lead extraction has been increasing in direct relationship to the increased numbers of cardiovascular implantable electronic devices. METHODS: Consecutive patients undergoing transvenous laser-assisted lead extraction at 13 centers were included. RESULTS: Between January 2004 and December 2007, 1,449 consecutive patients underwent laser-assisted lead extraction of 2,405 leads (20 to 270 procedures/site). Median implantation duration was 82.1 months (0.4 to 356.8 months). Leads were completely removed 96.5% of the time, with a 97.7% clinical success rate whereby clinical goals associated with the indication for lead removal were achieved. Failure to achieve clinical success was associated with body mass index 10 years and when performed in low volume centers. Major adverse events in 20 patients were directly related to the procedure (1.4%) including 4 deaths (0.28%). Major adverse effects were associated with patients with a body mass index or =2.0. Indicators of all-cause in-hospital mortality were pocket infections, device-related endocarditis, diabetes, and creatinine > or =2.0. CONCLUSIONS: Lead extraction employing laser sheaths is highly successful with a low procedural complication rate. Total mortality is substantially increased with pocket infections or device-related endocarditis, particularly in the setting of diabetes, renal insufficiency, or body mass index <25 kg/m(2). Centers with smaller case volumes tended to have a lower rate of successful extraction.
Wazni et al. (Mon,) conducted a cohort in Cardiovascular implantable electronic device lead extraction (n=1,449). Transvenous laser-assisted lead extraction was evaluated on Clinical success rate. Transvenous laser-assisted lead extraction achieved a 97.7% clinical success rate and 96.5% complete removal rate, with a 1.4% rate of procedure-related major adverse events.