PURPOSE: Detecting variations in myocardial water content with T2 mapping is superior to conventional T2 -weighted MRI since quantification enables direct observation of complicated pathology. Most commonly used T2 mapping techniques are limited in achievable spatial and/or temporal resolution, both of which reduce accuracy due to partial-volume averaging and misregistration between images. The goal of this study was to validate a novel free breathing T2 mapping sequence that overcomes these limitations. METHODS: The proposed technique was made insensitive to heart rate variability through the use of a saturation prepulse to reset magnetization every heartbeat. Respiratory navigator-gated, differentially T2 -weighted volumes were interleaved per heartbeat, guaranteeing registered images and robust voxel-by-voxel T2 maps. Free breathing acquisitions removed limits on spatial resolution and allowed short diastolic windows. Accuracy was quantified with simulations and phantoms. RESULTS: Homogeneous three-dimensional (3D) T2 maps were obtained from normal human subjects and swine. Normal human and swine left ventricular T2 values were 42.3 ± 4.0 and 43.5 ± 4.3 ms, respectively. The T2 value for edematous myocardium obtained from a swine model of acute myocardial infarction was 59.1 ± 7.1 ms. CONCLUSION: Free-breathing accurate 3D T2 mapping is feasible and may be applicable in myocardial assessment in lieu of current clinical black blood, T2 -weighted techniques.