Key points are not available for this paper at this time.
The LCModel method analyzes an in vivo spectrum as a Linear Combination of Model in vitro spectra from individual metabolite solutions. Complete model spectra, rather than individual resonances, are used in order to incorporate maximum prior information into the analysis. A nearly model-free constrained regularization method automatically accounts for the baseline and lineshape in vivo without imposing a restrictive parameterized form on them. LCModel is automatic (non-interactive) with no subjective input. Approximately maximum-likelihood estimates of the metabolite concentrations and their uncertainties (Cramér-Rao lower bounds) are obtained. LCModel analyses of spectra from users with fields from 1.5 to 9.4 T and a wide range of sequences, particularly with short TE, are used here to illustrate the capabilities and limitations of LCModel and proton MRS.
Stephen W. Provencher (Fri,) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: