ECG-gated CT revealed significant cardiac cycle-dependent changes in aortic annulus dimensions, with the systolic effective diameter being significantly larger than the diastolic diameter (mean difference 0.6 mm, p<0.001).
Observational (n=56)
ECG-gated CT demonstrates significant cardiac cycle dependency of aortic annulus dimensions, suggesting the systolic effective diameter is the most appropriate parameter for TAVI sizing to avoid paravalvular leaks.
Mean Difference: 0.6 (95% CI 0.2–0.9)
Absolute Event Rate: 25.8% vs 25.2%
p-value: p=<0.001
To evaluate the inter-individual variance and the variability of the aortic root dimensions during the cardiac cycle by computed tomography (CT) in patients with severe aortic stenosis prior to transcatheter aortic valve implantation (TAVI). Fifty-six patients (m/w = 16/40, 81 ± 6.8 years), scheduled for a transapical aortic valve implantation with available preprocedural ECG-gated CT were retrospectively included. The evaluation included sizing of the aortic annulus and the aortic sinus, measurements of the coronary topography, aortic valve planimetry and scoring of calcification. The new defined aortic annulus sphericity ratio revealed a mostly elliptical shape with increasing diastolic deformation. The calculated effective diameter (ED), determined from the annulus' lumen area, turned out to be the parameter least affected from cardiac cycle changes while systolic and diastolic annulus dimensions and shape (diameter and area) differed significantly (p < 0.001). In about 70 % of the patients with relevant paravalvular leaks the finally implanted prosthesis was too small according to the CT based calculated ED. The ostial height of the coronaries showed a high variability with a critical minimum range <5 mm. The degree of the aortic calcification did not have an influence on the aortic annulus deformation during the cardiac cycle, but on the occurrence of paravalvular leaks. The aortic root anatomy demonstrated a high inter-individual variability and cardiac cycle dependency. These results must be strongly considered during the patient evaluation prior to TAVI to avoid complications. The systolic effective diameter, as measured by ECG-gated CT, represents an appropriate parameter for sizing the aortic annulus.
Lehmkuhl et al. (Mon,) conducted a observational in Severe aortic stenosis (n=56). ECG-gated multi-slice computed tomography was evaluated on Effective diameter (ED) of the aortic annulus in systole versus diastole (MD 0.6, 95% CI 0.2-0.9, p=<0.001). ECG-gated CT revealed significant cardiac cycle-dependent changes in aortic annulus dimensions, with the systolic effective diameter being significantly larger than the diastolic diameter (mean difference 0.6 mm, p<0.001).