PURPOSE: To determine quantitative quality control procedures to evaluate technical variability in multi-center measurements of the diffusion coefficient of water as a prerequisite to use of the biomarker apparent diffusion coefficient (ADC) in multi-center clinical trials. MATERIALS AND METHODS: A uniform data acquisition protocol was developed and shared with 18 participating test sites along with a temperature-controlled diffusion phantom delivered to each site. Usable diffusion weighted imaging data of ice water at five b-values were collected on 35 clinical MRI systems from three vendors at two field strengths (1.5 and 3 Tesla T) and analyzed at a central processing site. RESULTS: Standard deviation of bore-center ADCs measured across 35 scanners was 10%) vendor-specific and system-specific spatial nonuniformity ADC bias was detected for the off-center measurement that was consistent with gradient nonlinearity. CONCLUSION: Standardization of DWI protocol has improved reproducibility of ADC measurements and allowed identifying spatial ADC nonuniformity as a source of error in multi-site clinical studies.