Key points are not available for this paper at this time.
We propose a non-contrast-enhanced, three-dimensional, free-breathing, electrocardiogram-gated, gradient recalled echo sequence with a slab-selective inversion for pulmonary vein (PV) MRI. A sagittal inversion slab was applied prior to data acquisition to suppress structures adjacent to the left atrium (LA) and PVs, thereby improving the conspicuity of the PV and LA. Compared with other MR angiography methods using an inversion pulse, the proposed method does not require signal subtraction and the inversion slab is not parallel to the imaging slab. The feasibility of the proposed method was demonstrated in healthy subjects. The inversion slab thickness and inversion time were optimized to be 60 mm and 500 ms, respectively. Compared to conventional gradient recalled echo imaging without inversion, the proposed technique significantly increased the contrast-to-noise ratios between the LA and the right atrium by 20-fold (P 0.3). The proposed technique greatly enhances the conspicuity of the PVs and LA without significant loss of signal-to-noise ratio.
Hu et al. (Sat,) studied this question.