Magnetic resonance imaging in 19 patients with abandoned cardiac device leads resulted in 0 adverse events within 7 days and no lead malfunctions upon generator reimplantation.
Observational (n=19)
No
Does magnetic resonance imaging cause adverse events or lead malfunction in patients with abandoned CIED leads?
MRI appears feasible and safe in patients with abandoned cardiac device leads when performed under careful monitoring, challenging the traditional contraindication.
INTRODUCTION: Abandoned cardiovascular implantable electronic device (CIED) leads remain a contraindication to magnetic resonance imaging (MRI) studies, largely due to in vitro data showing endocardial heating secondary to the radiofrequency field. We tested the hypothesis that abandoned CIED leads do not pose an increased risk of clinical harm for patients undergoing MRI. METHODS: This single-center retrospective study examined the outcomes of patients who had device generators removed before MRI, rendering the device leads abandoned. Information was gathered through chart review. Data collected included lead model, pacing threshold before MRI, anatomic region examined, threshold data after generator reimplantation, and clinical patient outcome. RESULTS: Patients (n = 19, 11 men and eight women) ranged in age from 19 to 85 at the time of MRI. There was a mean of 1.63 abandoned leads at the time of imaging; none of the leads were MRI conditional. Of the three implantable cardioverter defibrillator (ICD) leads, two of three were dual coil. Most (31/35) of the scans performed were of the central nervous system, including head and spinal imaging. There were no adverse events associated with MRI in any of these patients with abandoned leads within 7 days of the scan. No lead malfunctions or clinically significant change in pacing thresholds were noted with generator reimplantation. CONCLUSION: The use of MRI in patients with abandoned cardiac device leads appears feasible when performed under careful monitoring, with no adverse events, although the experience is small. MRI did not affect the function of leads that were subsequently reconnected to a cardiac device.
Higgins et al. (Fri,) conducted a observational in Abandoned pacemaker and defibrillator leads (n=19). Magnetic Resonance Imaging (MRI) was evaluated on Adverse events associated with MRI within 7 days of the scan. Magnetic resonance imaging in 19 patients with abandoned cardiac device leads resulted in 0 adverse events within 7 days and no lead malfunctions upon generator reimplantation.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: