What is the diagnostic accuracy of the AxSYM troponin I immunoassay for detecting acute myocardial infarction in patients presenting with chest pain?
The AxSYM troponin I immunoassay provides high diagnostic sensitivity and specificity for detecting acute myocardial infarction in patients with chest pain.
We evaluated the AxSYM troponin I (cTnI) immunoassay for assisting in the detection of acute myocardial infarction (AMI). At four sites, the total imprecision (CV) over 20 days was 6.3-10.2%. The minimum detectable concentration was 0.14 +/- 0.05 microgram/L. Comparison of cTnI measurements between the AxSYM and Stratus (n = 406) over the dynamic range of the AxSYM assay demonstrated good correlation, r = 0.881, with a proportional bias: AxSYM cTnI = 3.50(Stratus cTnI) - 1. 10. The confidence intervals (95%) for the slope and intercept were 3.39-3.64 and -1.32 to -0.95, respectively. The expected cTnI concentration in healthy individuals was /=96%, in skeletal muscle injury, chronic renal disease, and same-day noncardiac surgery patients.
Apple et al. (Mon,) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: