Do CARTO contact and EnSite noncontact mapping systems accurately identify ventricular scar compared to histology in a postinfarct ovine model?
Both CARTO contact and EnSite noncontact mapping systems demonstrate moderate accuracy in identifying postinfarct ventricular scar when validated against histology in an ovine model.
BACKGROUND: Substrate-based ablation is useful for nonhemodynamically tolerated postinfarct ventricular tachycardia. We assessed the accuracy of the CARTO contact and EnSite noncontact systems at identifying scar in a chronic ovine model with intramural plunge needle electrode recording and histological validation. METHODS AND RESULTS: Scar mapping was performed on 8 male sheep with previous percutaneous-induced myocardial infarction. Up to 20 plunge needles were inserted into the left ventricle of each animal in areas of dense scar, scar border, and normal myocardium. A simultaneous CARTO map and EnSite geometry were acquired using a single catheter, and needle electrode locations were registered. A dynamic substrate map was constructed using ratiometric 50% peak negative voltage. The scar percentage around each needle location was quantified histologically. Analysis was performed on 152 plunge needles and corresponding histological blocks. Spearman correlation with histology was 0.690 (P or =50% scar) was 0.896 for needle electrode PPV, 0.726 and 0.697 for CARTO bipolar and unipolar PPV, and 0.703 for EnSite dynamic substrate map (40 mm from the array were excluded.