The Arteriograph oscillometric method for assessing arterial stiffness showed highly significant correlations with SphygmoCor (r=0.67) and Complior (r=0.69) for pulse wave velocity (P<0.001).
Cross-Sectional (n=51)
51 patients who underwent arterial stiffness measurements using three different devices, with 35 patients repeating measurements after 1 week.
Arteriograph vs SphygmoCor and Complior
Correlation of pulse wave velocity (PWV) and augmentation index (AIx) — r = 0.67 (SphygmoCor) and r = 0.69 (Complior) for PWV; r = 0.92 for AIx, p=<0.001
Effect estimate: r = 0.67 (SphygmoCor) and r = 0.69 (Complior) for PWV; r = 0.92 for AIx
p-value: p=<0.001
INTRODUCTION: Pulse wave velocity (PWV) and augmentation index (AIx) are parameters of arterial stiffness and wave reflection. PWV and AIx are strong indicators for cardiovascular risk and are used increasingly in clinical practice. Previous systems for assessment of PWV and AIx are investigator dependent and time consuming. The aim of this study was to validate the new oscillometric method (Arteriograph) for determining PWV and AIx by comparing it to two clinically validated, broadly accepted tonometric and piezo-electronic systems (SphygmoCor and Complior). DESIGN AND METHOD: PWV and AIx were measured up to five times in 51 patients with the SphygmoCor, Complior and Arteriograph. In 35 patients, the measurements were repeated after 1 week in a second session using the same protocol. RESULTS: The correlations of the PWV as assessed with the Arteriograph with the values obtained using the SphygmoCor (r = 0.67, P < 0.001) and the Complior (r = 0.69, P < 0.001) were highly significant. Variability and reproducibility for PWV were best for the Arteriograph, followed by Complior and SphygmoCor. AIx (SphygmoCor versus Arteriograph) were very closely correlated (r = 0.92, P < 0.001). PERSPECTIVES: The Arteriograph is a new, easy-to-use and time-effective method for assessing arterial stiffness and wave reflection.
Building similarity graph...
Analyzing shared references across papers
Loading...
Johannes Baulmann
Orthopädische Praxis
Ulrich Schillings
University Hospital Bonn
Susanna Rickert
University Hospital Bonn
Journal of Hypertension
University of Bonn
GTx (United States)
University Hospital Bonn
Building similarity graph...
Analyzing shared references across papers
Loading...
Baulmann et al. (Sat,) reported a cross-sectional. Arteriograph vs. SphygmoCor and Complior was evaluated on Correlation of pulse wave velocity (PWV) and augmentation index (AIx) (r = 0.67 (SphygmoCor) and r = 0.69 (Complior) for PWV; r = 0.92 for AIx, p=<0.001). The Arteriograph oscillometric method for assessing arterial stiffness showed highly significant correlations with SphygmoCor (r=0.67) and Complior (r=0.69) for pulse wave velocity (P<0.001).