Does axial stent strut angle influence wall shear stress and risk of neointimal hyperplasia after stent implantation?
3D computational fluid dynamics models of stent foreshortening
Axial stent strut angle variation due to stent foreshortening
Wall shear stress and risk of neointimal hyperplasiasurrogate
Computational modeling suggests that stent foreshortening alters axial strut angle and wall shear stress, potentially increasing the risk of neointimal hyperplasia.
The current results suggest that foreshortening may predispose the stented vessel to a higher risk of neointimal hyperplasia.