OBJECTIVE: To determine the relationship between mean sensor glucose concentrations and hemoglobin A(1c) (HbA(1c)) values measured in the Diabetes Control and Complications Trial/Epidemiology of Diabetes Interventions and Complications laboratory at the University of Minnesota in a cohort of subjects with type 1 diabetes from the Juvenile Diabetes Research Foundation continuous glucose monitoring randomized trial. RESEARCH DESIGN AND METHODS: Near-continuous glucose sensor data (≥ 4 days/week) were collected for 3 months before a central laboratory-measured HbA(1c) was performed for 252 subjects aged 8-74 years, the majority of whom had stable HbA(1c) values (77% within ± 0.4% of the patient mean). RESULTS: The slope (95% CI) for mean sensor glucose concentration (area under the curve) versus a centrally measured HbA(1c) was 24.4 mg/dL (22.0-26.7) for each 1% change in HbA(1c), with an intercept of -16.2 mg/dL (-32.9 to 0.6). Although the slope did not vary with age or sex, there was substantial individual variability, with mean sensor glucose concentrations ranging from 128 to 187 mg/dL for an HbA(1c) of 6.9-7.1%. The root mean square of the errors between the actual mean sensor glucose concentration versus the value calculated using the regression equation was 14.3 mg/dL, whereas the median absolute difference was 10.1 mg/dL. CONCLUSIONS: There is substantial individual variability between the measured versus calculated mean glucose concentrations. Consequently, estimated average glucose concentrations calculated from measured HbA(1c) values should be used with caution.