A newly developed ultrasound method accurately quantified calcium amounts in an in vitro model, demonstrating a strong correlation with computed tomography calcium scores (R2 = 0.7391, p<0.0001).
Does a new ultrasound method accurately quantify the amount of calcium compared to computed tomography in an in vitro phantom model?
A novel ultrasound-based method accurately quantifies calcium amounts in an in vitro phantom model, showing strong correlation with micro-CT, suggesting potential for future clinical application in assessing valvular calcification.
Effect estimate: R2 = 0.7391
p-value: p=<0.0001
PURPOSE: Calcification is an important prognostic factor in aortic valve stenosis. However, there is no ultrasound (US) method available to accurately quantify calcification in this setting to date. We aimed to validate a new US method for measuring the amount of calcium in an in vitro model, and compare it to computed tomography (CT), the current imaging gold standard. MATERIALS AND METHODS: An agar phantom (2% agar) was made, containing 9 different amounts of calcium-hydroxyapatite Ca5(PO4)3OH (2 to 50 mg). The phantoms were imaged with micro-CT and US (10 MHz probe). The calcium area (areacalcium) and its maximum pixel value (PVmax) were obtained. These values were summed to calculate CT and US calcium scores (∑(areacalcium × PVmax)) and volumes (∑areacalcium). Both US- and CT-calcium scores were compared with the calcium amounts, and with each other. RESULTS: Both calcium scores correlated significantly with the calcium amount (R2 = 0.9788, p<0.0001 and R2 = 0.8154, p<0.0001 for CT and US respectively). Furthermore, there was a significant correlation between US and CT for calcium volumes (R2 = 0.7392, p<0.0001) and scores (R2 = 0.7391, p<0.0001). CONCLUSION: We developed a new US method that accurately quantifies the amount of calcium in an in vitro model. Moreover it is strongly correlated with CT.
Gillis et al. (Tue,) conducted a other in Calcific aortic valve stenosis (in vitro model) (n=9). Ultrasound (US) calcium scoring vs. Micro-CT calcium scoring was evaluated on Correlation between US calcium score and CT calcium score (R2 = 0.7391, p=<0.0001). A newly developed ultrasound method accurately quantified calcium amounts in an in vitro model, demonstrating a strong correlation with computed tomography calcium scores (R2 = 0.7391, p<0.0001).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: