Key points are not available for this paper at this time.
Formaldehyde (FA) is a potential breath marker for lung cancer and a tracer for indoor air quality monitoring. Its typical concentrations are below 100 ppb posing a sensitivity and selectivity challenge to current portable sensor systems. Here, we present a highly sensitive, selective, and compact electronic nose (E-nose) for real-time quantification of FA at realistic conditions. This E-nose consists of four nanostructured and highly porous Pt-, Si-, Pd-, and Ti-doped SnO2 sensing films directly deposited onto silicon wafer-based microsubstrates by flame spray pyrolysis (FSP). The constituent sensors offer stable responses (24 h tested) and detection of FA down to 3 ppb (signal-to-noise ratio > 25) at breath-realistic 90% relative humidity. Each dopant induces different analyte selectivity enabling selective detection of FA in two-, three- and four-analyte mixtures by multivariate linear regression. In simulated breath (FA with higher acetone, NH3, and ethanol concentrations), FA is detected with an average error ≤ 9 ppb using the present E-nose and overcoming selectivity issues of single sensors. This device could facilitate easy screening of lung cancer patients and monitoring of indoor FA concentrations.
Güntner et al. (2016) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: