Multidetector-row computed tomography accurately measured mechanical prosthetic heart valve leaflet motion compared with fluoroscopy, showing high agreement (ICC >0.989) and a maximal median discrepancy of 2 degrees.
Mechanical prosthetic heart valves (n=4)
Multidetector-row computed tomography (MDCT) vs Fluoroscopy
Leaflet closure angle measurements — ICC >0.989
Effect estimate: ICC >0.989
PURPOSE: The purpose of this study was to compare multidetector-row computed tomography (MDCT) leaflet restriction measurements with fluoroscopy measurements in commonly used mechanical prosthetic heart valves (PHVs). METHODS: Four mechanical PHVs (ON-X, Carbomedics, St. Jude, and Medtronic Hall) were imaged in a pulsatile model using fluoroscopy and 64-detector-row computed tomography. Five image acquisitions of each PHV without (1) and with (4) restricted leaflet closure were made. Three observers measured closure angles on fluoroscopy and MDCT. Data were analyzed using intraclass correlation coefficient (ICC) and Bland-Altman plots. RESULTS: Interobserver agreement was high in restricted and non-restricted leaflets on both modalities (ICCs >0.995). MDCT and fluoroscopy showed high agreements (ICCs >0.989). Median MDCT closure angle measurements differed at most -2 to +2 degrees from fluoroscopy in the restricted and -1 to +2 degrees in the non-restricted leaflets. CONCLUSIONS: MDCT allows measurement of leaflet motion with a maximal median discrepancy of 2 degrees. Both MDCT and fluoroscopy detect restricted leaflet closure with great accuracy.
Building similarity graph...
Analyzing shared references across papers
Loading...
Dominika Suchá
Utrecht University
Petr Symersky
University of British Columbia
Evert‐Jan Vonken
Heidelberg University
Journal of Computer Assisted Tomography
University Medical Center Utrecht
Amsterdam UMC Location University of Amsterdam
Building similarity graph...
Analyzing shared references across papers
Loading...
Suchá et al. (Wed,) conducted a other in Mechanical prosthetic heart valves (n=4). Multidetector-row computed tomography (MDCT) vs. Fluoroscopy was evaluated on Leaflet closure angle measurements (ICC >0.989). Multidetector-row computed tomography accurately measured mechanical prosthetic heart valve leaflet motion compared with fluoroscopy, showing high agreement (ICC >0.989) and a maximal median discrepancy of 2 degrees.
synapsesocial.com/papers/6a14e584cfc8ed0661c1ac9f — DOI: https://doi.org/10.1097/rct.0b013e3182ab5f15