Key points are not available for this paper at this time.
A fast swept-source optical coherence tomography (SS-OCT) system is employed to acquire volumes of dental tissue, in order to monitor the temporal evolution of dental wear. An imaging method is developed to evaluate the volume of tissue lost in ex vivo artificially induced abfractions and attritions. The minimal volume (measured in air) that our system could measure is 2352 μm3. A volume of 25,000 A-scans is collected in 2.5 s. All these recommend the SS-OCT method as a valuable tool for dynamic evaluation of the abfraction and attrition with remarkable potential for clinical use.
Uduak B. Ubom (Wed,) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: