Key points are not available for this paper at this time.
Size-exclusion chromatography coupled with SAXS (small-angle X-ray scattering), often performed using a flow-through capillary, should allow direct collection of monodisperse sample data. However, capillary fouling issues and non-baseline-resolved peaks can hamper its efficacy. The UltraScan solution modeler ( US-SOMO ) HPLC-SAXS (high-performance liquid chromatography coupled with SAXS) module provides a comprehensive framework to analyze such data, starting with a simple linear baseline correction and symmetrical Gaussian decomposition tools Brookes, Pérez, Cardinali, Profumo, Vachette & Rocco (2013). J. Appl. Cryst. 46 , 1823–1833. In addition to several new features, substantial improvements to both routines have now been implemented, comprising the evaluation of outcomes by advanced statistical tools. The novel integral baseline-correction procedure is based on the more sound assumption that the effect of capillary fouling on scattering increases monotonically with the intensity scattered by the material within the X-ray beam. Overlapping peaks, often skewed because of sample interaction with the column matrix, can now be accurately decomposed using non-symmetrical modified Gaussian functions. As an example, the case of a polydisperse solution of aldolase is analyzed: from heavily convoluted peaks, individual SAXS profiles of tetramers, octamers and dodecamers are extracted and reliably modeled.
Building similarity graph...
Analyzing shared references across papers
Loading...
Journal of Applied Crystallography
Centre National de la Recherche Scientifique
Commissariat à l'Énergie Atomique et aux Énergies Alternatives
Université Paris-Sud
Add This Paper to Your Research Feed
Any time a new paper drops it will be there.
Brookes et al. (Thu,) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: