Arm oscillometric devices achieved high reliability for systolic BP in atrial fibrillation (ICC 0.94 for Omron HEM907), but agreement with invasive measurement was low unless averaged over 3 readings.
Cross-Sectional (n=58)
Randomized sequences
Do oscillometric blood pressure devices provide reliable and accurate measurements compared to invasive arterial blood pressure in patients with atrial fibrillation?
Arm-based oscillometric devices show high reliability for systolic blood pressure in atrial fibrillation, and agreement with invasive measurements improves when averaging three consecutive readings.
Effect estimate: ICC 0.94 (95% CI 0.90-0.97)
BACKGROUND: The use of automated (oscillometric) blood pressure (BP) devices is not validated in atrial fibrillation (AF) patients. OBJECTIVES: To assess the reliability of three oscillometric BP devices, and the agreement with invasive arterial blood pressure(IBP) in AF patients. METHODS: 48 AF patients with randomized sequences of 10 consecutive BP measurements with two pairs of devices: (1) OmronR7™(wrist) and OmronHEM907™(arm); (2) OmronR7™ and Microlife WatchBPhome(arm). Reliability and agreement of each device were assessed by the intra-class correlation coefficient (ICC) for the continuous BP measurements and Bland 0.97] vs. 0.77 0.67; 0.89; Microlife WatchBPhome:0.92 0.88; 0.96 vs.0.79 0.69; 0.89).The correlation coefficient between Microlife WatchBPhome and IBP computed using the average of repeated measurements from two to ten measurements improved up to the third and remained stable afterwards. The agreement between IBP and SBP, and IBP and DBP, was moderate as illustrated by a wide limit of agreement -24; 26(SBP) and -15;17(DBP) for Microlife WatchBPHome, respectively and -30; 13(SBP) and -7; 15(DBP) for OmronHEM907. CONCLUSIONS: BP measurement using the two arm oscillometric devices achieved a high reliability for SBP. The agreement between IBP and arm devices was low but using the average of three consecutive measurements improved the results substantially.
Halfon et al. (Thu,) conducted a cross-sectional in Atrial fibrillation (n=58). Arm oscillometric BP devices (Omron HEM907 and Microlife WatchBPhome) vs. Invasive arterial blood pressure (IBP) was evaluated on Reliability (intra-class correlation coefficient) for systolic blood pressure (ICC 0.94, 95% CI 0.90-0.97). Arm oscillometric devices achieved high reliability for systolic BP in atrial fibrillation (ICC 0.94 for Omron HEM907), but agreement with invasive measurement was low unless averaged over 3 readings.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: