Does the Persyst 14 algorithm perform noninferiorly to expert humans in detecting seizures on prolonged EEG recordings?
The Persyst 14 seizure detection algorithm performs statistically noninferior to human experts in evaluating prolonged EEG recordings.
PURPOSE: To compare the seizure detection performance of three expert humans and two computer algorithms in a large set of epilepsy monitoring unit EEG recordings. METHODS: One hundred twenty prolonged EEGs, 100 containing clinically reported EEG-evident seizures, were evaluated. Seizures were marked by the experts and algorithms. Pairwise sensitivity and false-positive rates were calculated for each human-human and algorithm-human pair. Differences in human pairwise performance were calculated and compared with the range of algorithm versus human performance differences as a type of statistical modified Turing test. RESULTS: A total of 411 individual seizure events were marked by the experts in 2,805 hours of EEG. Mean, pairwise human sensitivities and false-positive rates were 84.9%, 73.7%, and 72.5%, and 1.0, 0.4, and 1.0/day, respectively. Only the Persyst 14 algorithm was comparable with humans-78.2% and 1.0/day. Evaluation of pairwise differences in sensitivity and false-positive rate demonstrated that Persyst 14 met statistical noninferiority criteria compared with the expert humans. CONCLUSIONS: Evaluating typical prolonged EEG recordings, human experts had a modest level of agreement in seizure marking and low false-positive rates. The Persyst 14 algorithm was statistically noninferior to the humans. For the first time, a seizure detection algorithm and human experts performed similarly.