The YuWell YE900 blood pressure monitor met AAMI/ESH/ISO validation criteria, with a mean systolic/diastolic difference of 1.7 ± 6.62 / 3.1 ± 5.76 mmHg compared to reference readings.
Cross-Sectional (n=85)
Does the YuWell YE900 blood pressure monitor accurately measure blood pressure compared to reference standards in adults and children?
The YuWell YE900 professional electronic blood pressure monitor meets the AAMI/ESH/ISO Universal Standard requirements for accuracy in adults and children and can be recommended for clinical use.
Mean Difference: 1.7
OBJECTIVE: To evaluate the accuracy of the YuWell YE900 oscillometric upper-arm professional office blood pressure monitor in adults and children according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard (ISO 81060-2:2018). METHODS: Subjects were recruited to fulfill the age, sex, blood pressure and cuff distribution criteria of the AAMI/ESH/ISO Universal Standard in adults and children (aged 3-12 years) using the same arm sequential blood pressure measurement method. Three cuffs of the test device were used for arm circumference 18-22 cm (small), 22-32 cm (medium) and 32-42 cm (large). RESULTS: Ninety-two subjects were recruited, and 85 (50 adults and 35 children) were analyzed. For validation criterion 1, the mean ± SD of the differences between the test device and reference blood pressure readings was 1.7 ± 6.62/3.1 ± 5.76 mmHg (systolic/diastolic). For validation criterion 2, the SD of the averaged blood pressure differences between the test device and reference blood pressure per subject was 5.25/5.13 mmHg (systolic/diastolic). CONCLUSION: The YuWell YE900 professional electronic blood pressure monitor has passed the requirements of the AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018) in adults and children and can be recommended for clinical use.
Zhang et al. (Mon,) reported a cross-sectional. YuWell YE900 oscillometric blood pressure monitor vs. Reference blood pressure readings was evaluated on Mean difference between the test device and reference blood pressure readings (validation criterion 1) (Mean ± SD difference 1.7 ± 6.62 mmHg (systolic) and 3.1 ± 5.76 mmHg (diastolic)). The YuWell YE900 blood pressure monitor met AAMI/ESH/ISO validation criteria, with a mean systolic/diastolic difference of 1.7 ± 6.62 / 3.1 ± 5.76 mmHg compared to reference readings.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: