The HeartGuide wearable blood pressure monitor reliably detected systolic BP decreases during tilt table testing compared to continuous finger monitoring (Lin's CCC=0.78, P=0.001).
Observational (n=81)
Does a wearable watch-type blood pressure monitor reliably measure systolic blood pressure drops compared to finger beat-to-beat monitoring during impending syncopal episodes in patients evaluated for reflex syncope?
A wearable watch-type blood pressure monitor can feasibly and reliably detect systolic blood pressure drops during presyncope induced by tilt table testing.
Effect estimate: Lin's concordance correlation coefficient=0.78
Absolute Event Rate: -28.5% vs -30.3%
p-value: p=0.001
Background We assessed the reliability and feasibility of blood pressure (BP) measurements by means of a new wearable watch‐type BP monitor (HeartGuide) in detecting episodes of hypotensive (pre)syncope induced by tilt table test. Methods and Results An intrapatient comparison between systolic BP (SBP) measured by means of the HeartGuide device and noninvasive finger beat‐to‐beat BP monitoring was undertaken both at baseline in supine position and repeatedly during tilt table test in patients evaluated for reflex syncope. Intrapatient fall of systolic BP from baseline was measured. Eighty‐one patients (mean age, 61±19 years; 46 women) were included. Overall, HeartGuide was able to yield BP values at the time of BP nadir in 58 (72%) patients (average HeartGuide SBP 102±18 mm Hg, versus finger SBP 101±19 mm Hg). Compared with baseline, the maximum SBP decrease was on average −28.5±27.8 and −30.3±33.9 mm Hg respectively (Lin's concordance correlation coefficient=0.78, r =0.79, P =0.001). In the subgroup of 38 patients with tilt table test induced (pre)syncope, the average HeartGuide SBP during symptoms was 97±16 mm Hg, and the finger SBP was 94±18 mm Hg. Compared with baseline, the maximum SBP decrease was on average −35.2±29.3 and −43.3±31.8 mm Hg, respectively (Lin's concordance correlation coefficient=0.83, r =0.87, P =0.001). Conclusions Our data indicate that the HeartGuide BP monitor can detect low BP during presyncope and that its measure of SBP change is consistent with that simultaneously obtained through continuous BP monitoring, despite some intrapatient variability. Thus, this device might be useful in determining the hypotensive nature of spontaneous (pre)syncopal symptoms, a possibility that should be verified by field studies.
Groppelli et al. (Fri,) conducted a observational in reflex syncope (n=81). Wearable watch-type BP monitor (HeartGuide) vs. Noninvasive finger beat-to-beat BP monitoring was evaluated on Intrapatient fall of systolic BP from baseline (Lin's concordance correlation coefficient=0.78, p=0.001). The HeartGuide wearable blood pressure monitor reliably detected systolic BP decreases during tilt table testing compared to continuous finger monitoring (Lin's CCC=0.78, P=0.001).