Key points are not available for this paper at this time.
To assess the analytic quality of laboratory testing in the United States, we obtained proficiency testing survey results from several national programs that comply with Clinical Laboratory Improvement Amendments (CLIA) regulations. We studied regulated tests (cholesterol, glucose, calcium, fibrinogen, and prothrombin time) and nonregulated tests (international normalized ratio INR, glycohemoglobin, and prostate-specific antigen PSA). Quality was assessed on the σ scale with a benchmark for minimum process performance of 3 σ and a goal for world-class quality of 6 σ. Based on the CLIA criteria for acceptable performance in proficiency testing (allowable total errors TEa), the national quality of cholesterol testing values as 2.9 to 3.0; glucose (TEa = 10%) estimated σ (TEa = 10%), 2.9 to 3.3; calcium (TEa = 1.0 mg/dL), 2.8 to 3.0; prothrombin time (TEa = 15%), 1.8; INR (TEa = 20%), 2.4 to 3.5; fibrinogen (TEa = 20%), 1.8 to 3.2; glycohemoglobin (TEa = 10%), 1.9 to 2.6; and PSA (TEa = 10%), 1.2 to 1.8. The analytic quality of laboratory tests requires improvement in measurement performance and more intensive quality control monitoring than the CLIA minimum of 2 levels per day.
Westgard et al. (Wed,) studied this question.