Key points are not available for this paper at this time.
Single ion counting in particle therapy may lead to new beam monitoring systems, enabling innovative delivery strategies that are faster and more sensitive than those currently used in clinics. Previous studies carried out by the University and the National Institute of Nuclear Physics (INFN) of Turin have demonstrated the feasibility of using thin silicon detectors to count single protons in clinical beams (Monaco et al., 2023) 1. The aim of this work is to report the performance of a strip-segmented 60-μm thick silicon PIN sensor used for single carbon ion discrimination. All measurements were performed using the CNAO synchrotron at different beam energies covering the clinical energy range (115–399 MeV/u). Signals from the sensor strips were read using a custom amplifier board and sampled with a 5 GS/s digitizer. The carbon ion signals were analyzed in terms of amplitude, duration, and deposited charge at different sensor bias voltages.
Building similarity graph...
Analyzing shared references across papers
Loading...
Diango Montalvan Olivares
O.A. Martì Villarreal
M. Abujami
Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment
University of Turin
Istituto Nazionale di Fisica Nucleare, Sezione di Torino
Fondazione Bruno Kessler
Building similarity graph...
Analyzing shared references across papers
Loading...
Olivares et al. (Mon,) studied this question.
www.synapsesocial.com/papers/68e7285cb6db6435876a206a — DOI: https://doi.org/10.1016/j.nima.2024.169289