Abstract We aimed to evaluate the accuracy and repeatability of the T 1, T 2*, and proton density (PD) values obtained by quantitative parameter mapping (QPM) using the ISMRM/NIST MRI system phantom and compared them with computer simulations. We compared the relaxation times and PD obtained through QPM with the reference values of the ISMRM/NIST MRI system phantom and conventional methods. Furthermore, we evaluated the presence or absence of influences other than noise in T 1 and T 2* values obtained by QPM by comparing the obtained coefficient of variation (CV) with simulation results. The T 1, T 2*, and PD values by QPM showed a strong correlation with the measured values and the referenced values. The simulated CVs of QPM calculated for each sphere showed similar trends to those of the actual scans.