Key points are not available for this paper at this time.
Abstract Background There is limited evidence demonstrating the ability of audiometric bone vibrators to elicit ocular vestibular evoked myogenic potentials (oVEMPs). The RadioEar B71 bone vibrator has insufficient power output to reliably evoke oVEMPs, which has previously left nonaudiometric and nonmedically approved devices such as the Brüel 22–47 years) with no history of hearing, balance, or neurological disorders were included in the study. Data collection and analysis One-cycle alternating tone-burst stimuli at 500 Hz were delivered to the mastoid from each transducer. The vibratory force levels were matched at 127 dB peak-to-peak equivalent force levels, and both were held in place with a static force around 10 N. oVEMP waveforms were gathered from the contralateral eye using the belly-tendon montage and were assessed for statistical equivalence. Results There was an absence of any statistically significant difference in N10 and N10–P15 amplitudes in oVEMPs from each transducer. Conclusions Our results indicate that B81 can elicit oVEMPs with no meaningful differences to those from the Mini-shaker, provided effective stimulus levels are matched and static force is sufficient. Although further work is necessary to investigate equivalence at other stimulus frequencies and stimulation sites, the results support the use of the B81 to elicit 500Hz oVEMPs at the mastoid in a clinical setting.
Garbutt et al. (2024) studied this question.