The BHx streak tube, under development at the Laboratory for Laser Energetics, incorporates a series of novel electron-optics elements to enable high fidelity measurements for high-energy-density physics experiments. The system is engineered to support a 25-mm active photocathode region compatible with sub-picosecond temporal resolution and with 70% internal photoelectron throughput. It exhibits negligible geometric distortion on a flat output screen, making it well suited for variety of measurements, such as ultrafast x-ray spectroscopy. This paper presents characterization data from a prototype unit tested with an ultraviolet laser in a static deflection (non-swept) mode and shows good agreement with the predictions from numerical modeling, including focusing performance and geometric distortion measurements. Key design elements have been demonstrated and de-risked, laying the foundation for dynamic deflection testing. The anticipated improvements in data fidelity are expected to impact the fields as diverse as inertial confinement fusion, laboratory astrophysics, and materials science.