Abstract Objectives The objective of this study is to evaluate the analytical and diagnostic performance of a high-sensitivity point-of-care (POC) cardiac troponin I assay, the Quidel TriageTrue™ (QuidelOrtho Inc, San Diego, USA), compared to central laboratory testing (CLT) in accelerated diagnostic protocols (ADP) in real time in a clinical environment. Methods In a nested sub-study of a pragmatic randomised control trial, consecutive patients with suspected acute coronary syndrome (ACS) and chest pain 50 % lower than 99th percentile 20.5 ng/L) and a 20 % CV at 1.2 ng/L. Receiver operator characteristics (ROC) curves were computed for each assay against adjudicated index type 1 MI to study clinical performance. At all-time points there were excellent performance for whole blood POC TriageTrue: area under the curve (AUC) 0.97 95 % CI 0.94–098, 0.98 95 % CI 0.97–1.00 and 0.95 95 % CI 0.92–0.98 at time 0, 1 and 3 h respectively. There was statistical equivalence for performance of whole blood and plasma POC TriageTrue hs-TnI and laboratory Siemens Atellica hs-TnI. Conclusions The whole blood POC TriageTrue hs-TnI assay demonstrates imprecision levels consistent with high sensitivity characteristics and has a clinical performance equivalent to an established, validated and NICE approved laboratory Siemens Atellica hs-TnI.
Dakshi et al. (Mon,) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: