Does the CareUp smartwatch accurately estimate blood pressure compared to standard sphygmomanometers and oscillometry-based devices in human subjects?
The CareUp smartwatch provides a potentially accurate, wearable method for real-time blood pressure monitoring using photoplethysmography.
We present a novel smartwatch, CareUp ® , for estimating the Blood Pressure (BP) in real time. It consists of two pulse oximeters: one placed on the back and one on the front of the device. Placing the index finger on the front oximeter starts the acquisition of two photoplethysmograms (PPG); the signals are then filtered and cross-correlated to obtain a Time Delay between them, called Pulse Transit Time (PTT). The Heart Rate (HR) (estimated from the finger PPG) and the PTT are then input in a linear model to give an estimation of the Systolic and Diastolic BP. The performance of the smartwatch in measuring BP have been validated in the Institut Coeur Paris Centre Turin (ICPC), using a sphygmomanometer, on 44 subjects. During the validation, the measures of the CareUp ® were compared to those of two oscillometry-based devices already available on the market: Thuasne ® and Magnien ® . The results showed an accuracy comparable to the oscillometry-based devices and they almost agreed with the American Association for the Advancement of Medical Instrumentation standard for non-automated sphygmomanometers. The integration of the BP estimation algorithm in the smartwatch makes the CareUp ® an easy-to-use, wearable device for monitoring the BP in real time.
Lazazzera et al. (Tue,) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: