A Comparison of the Vasotrac with Invasive Arterial Blood Pressure Monitoring in Children After Pediatric Cardiac Surgery

View Full Paper
LCL. CliffordKTKristi ThomasDZDavid Zurakowski

Key Result

Noninvasive arterial blood pressure monitoring using the Vasotrac agreed well with invasive arterial catheters in pediatric patients (absolute mean difference for mean BP 3.5 mm Hg; 95% CI 2.5-4.0).

Study Design

Type

Observational (n=16)

Structured PICO

Does the Vasotrac device provide accurate arterial blood pressure measurements compared to an arterial catheter in children after pediatric cardiac surgery?

P
Population
16 children (mean age 10.1 years) undergoing corrective cardiac surgery who had simultaneous noninvasive and invasive arterial blood pressure measurements.
E
Exposure
Vasotrac device for near-continuous noninvasive arterial blood pressure monitoring
C
Comparator
Invasive arterial catheter monitoring
O
Outcome
Agreement between measurements of arterial blood pressure (systolic, diastolic, mean) and heart rate obtained from the Vasotrac versus an arterial cathetersurrogate

The Vasotrac device provides noninvasive arterial blood pressure measurements that agree well with invasive arterial monitoring in pediatric patients after cardiac surgery.

Main Result

Mean Difference: 3.5 (95% CI 2.5–4)

p-value: p=<0.001

Abstract

The Vasotrac is a device that provides near-continuous and noninvasive arterial blood pressure monitoring and may be an alternative to direct intraarterial measurement. It has been evaluated in adult patients, but minimal information is available for pediatric patients. We evaluated agreement between measurements of arterial blood pressure and heart rate obtained from the Vasotrac versus an arterial catheter in a pediatric population. Children undergoing corrective cardiac surgery were enrolled. Simultaneous arterial blood pressure measurements were obtained postoperatively from the Vasotrac unit and an arterial catheter. Bland-Altman plots were constructed to assess agreement. Paired correlation analysis, bias, and precision calculations were performed. Sixteen patients, mean age 10.1 +/- 2.3 yr and weight 34.6 +/- 11.9 kg, were enrolled. Four-thousand-one- hundred- two paired measurements were obtained. Arterial blood pressures measured noninvasively correlated with catheter measurements with Pearson r values of 0.90, 0.80, and 0.91 for systolic, diastolic, and mean arterial blood pressures, respectively (all P < 0.001). There was excellent agreement between arterial blood pressure measurement methods. Absolute mean differences based on mixed-model regression with 95% confidence intervals were 4.0 mm Hg (3.0-5.0 mm Hg), 4.3 mm Hg (3.1-5.5 mm Hg), and 3.5 mm Hg (2.5-4.0 mm Hg) for systolic blood pressure, diastolic blood pressure, and mean blood pressure, respectively. Arterial blood pressure measurements obtained from the Vasotrac agreed well with invasive arterial monitoring in pediatric patients.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Clifford et al. (2005) conducted an observational in Pediatric cardiac surgery (n=16). Vasotrac noninvasive arterial blood pressure monitoring vs. Invasive arterial catheter was evaluated on Absolute mean difference in mean arterial blood pressure (MD 3.5 mm Hg, 95% CI 2.5-4.0, p=<0.001). Noninvasive arterial blood pressure monitoring using the Vasotrac agreed well with invasive arterial catheters in pediatric patients (absolute mean difference for mean BP 3.5 mm Hg; 95% CI 2.5-4.0).

synapsesocial.com/papers/6a1fe94275fc4a116b2e4b47https://doi.org/10.1213/01.ane.0000149591.67207.05
Ask AI
Helpful
Bookmark
Share
View Full Paper

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Blood Pressure Measurement by Arterial Tonometry in Controlled Hypotension1991 · 90 citations
  2. 2A New Noninvasive Method to Measure Blood Pressure1999 · 103 citations
  3. 3STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT1986 · 48,054 citations
  4. 4Tissue hypoxia distal to a Peñaz finger blood pressure cuff1985 · 43 citations
  5. 5COMPARISON OF AUTOMATIC OSCILLOMETRIC ARTERIAL PRESSURE MEASUREMENT WITH CONVENTIONAL AUSCULTATORY MEASUREMENT IN THE LABOUR WARD1993 · 34 citations