October 2, 2025ACS Sensors3 citationsOpen Access

A Printable OECT for Simple Integration in Nitrocellulose-Based Assays

View Full Paper
MCMartina CicoliniASAli SolgiLVLorenzo Vigna

Key Points

  • The printable OECT enables quantitative biochemical sensing, improving sensitivity at point-of-care applications.
  • Achieving a limit of detection of 0.01 mM for dopamine, the system enhances reliability over traditional outputs in LFAs.
  • This microfluidic platform integrates seamlessly with cellulose membranes, ensuring contamination-free operations in diagnostics.
  • The maximum transconductance observed was approximately 4 mS, suggesting robust performance in real-time biochemical monitoring.

Abstract

Paper-based biosensors hold significant promise for point-of-care (POC) diagnostic applications. Among these, lateral flow assays (LFAs) are particularly appealing due to their ease of use, portability, and low cost. However, their limited sensitivity and qualitative output set drawbacks on their reliability and widespread application. In response to the growing need for rapid and consistent diagnostic and monitoring tools, Organic Electrochemical Transistors (OECTs) have emerged as powerful devices in biochemical sensing applications because of their high sensitivity, low operating voltage, and compatibility with a biological environment. In this work, we developed a printable OECT for biochemical sensing on a commercial cellulose membrane, commonly used as a detection substrate in LFA-based rapid tests. Constituting a self-standing, passive microfluidic platform, the system was designed to transport and interact with liquid samples, while ensuring a contamination-free zone for the active components. Inside a dry area delimited by a hydrophobic barrier, the OECT components include dispense-printed silver electrodes, a polystyrenesulfonate-doped poly(3,4-ethylenedioxy-thiophene) (PEDOT:PSS) channel and gate, and a solid-state electrolyte (SSE) layer. Outside the dry area, a PEDOT:PSS extended gate alone interacts with the analyte in the liquid sample, preventing channel contamination and enhancing the system stability. We investigated the effect of dopamine (DA) oxidation at the extended gate interface on the device response and observed variations in the transfer characteristics, transconductance and Ion/Ioff ratio, obtaining a limit of detection of 0.01 mM. With a maximum transconductance of approximately 4 mS, our system shows potential for the integration of an easy-to-fabricate device into an affordable biochemical assay, providing quantitative results at the point-of-care site to complement and reinforce the typical colorimetric response of LFAs.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Cicolini et al. (2025) studied this question.

synapsesocial.com/papers/68de8eaeaa6cec72c69ea817https://doi.org/10.1021/acssensors.5c01893
Ask AI
Helpful
Bookmark
Share
View Full Paper

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Three-Dimensional Printing of Highly Conducting PEDOT: PSS-Based Polymers2022 · 20 citations
  2. 2New trends in the electrochemical sensing of dopamine2012 · 464 citations
  3. 3Comparison of seven commercial SARS-CoV-2 rapid point-of-care antigen tests: a single-centre laboratory evaluation study2021 · 342 citations
  4. 4Point‐Of‐Care Ultra‐Portable Single‐Molecule Bioassays for One‐Health2023 · 65 citations
  5. 5All-printed and stretchable organic electrochemical transistors using a hydrogel electrolyte2023 · 52 citations