PulseExploreJournal ClubDebatesTrendingResearchersJournals
Instagram
HomeExploreJournal ClubTrending
Synapse
⌘+K
Synapse
June 21, 20240 citationsOpen Access

CCAT: Detector Noise Limited Performance of the RFSoC-based Readout Electronics for mm/sub-mm/far-IR KIDs

View Full Paper
ASAdrian K. SinclairJBJames BurgoyneAHAnthony I. Huber

Key Points

Key points are not available for this paper at this time.

Abstract

The Fred Young Submillimeter Telescope (FYST), on Cerro Chajnantor in the Atacama desert of Chile, will conduct wide-field and small deep-field surveys of the sky with more than 100,000 detectors on the Prime-Cam instrument. Kinetic inductance detectors (KIDs) were chosen as the primary sensor technology for their high density focal plane packing. Additionally, they benefit from low cost, ease of fabrication, and simplified cryogenic readout, which are all beneficial for successful deployment at scale. The cryogenic multiplexing complexity is pulled out of the cryostat and is instead pushed into the digital signal processing of the room temperature electronics. Using the Xilinx Radio Frequency System on a Chip (RFSoC), a highly multiplexed KID readout was developed for the first light Prime-Cam and commissioning Mod-Cam instruments. We report on the performance of the RFSoC-based readout with multiple detector arrays in various cryogenic setups. Specifically we demonstrate detector noise limited performance of the RFSoC-based readout under the expected optical loading conditions.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Sinclair et al. (2024) studied this question.

synapsesocial.com/papers/68e63e20b6db6435875cf9eahttps://doi.org/10.48550/arxiv.2406.14892
Ask AI
Helpful
Bookmark
Share
View Full Paper