PulseExploreJournal ClubDebatesTrendingResearchersJournals
Instagram
HomeExploreJournal ClubTrending
Synapse
⌘+K
Synapse
March 1, 2024The Bone & Joint Journal8 citations

Accuracy of femoral component anteversion in robotic total hip arthroplasty

View Full Paper
NSNobuhiko SuganoYMYuki MaedaHFH. Fuji

Key Points

Key points are not available for this paper at this time.

Abstract

Aims Femoral component anteversion is an important factor in the success of total hip arthroplasty (THA). This retrospective study aimed to investigate the accuracy of femoral component anteversion with the Mako THA system and software using the Exeter cemented femoral component, compared to the Accolade II cementless femoral component. Methods We reviewed the data of 30 hips from 24 patients who underwent THA using the posterior approach with Exeter femoral components, and 30 hips from 24 patients with Accolade II components. Both groups did not differ significantly in age, sex, BMI, bone quality, or disease. Two weeks postoperatively, CT images were obtained to measure acetabular and femoral component anteversion. Results The mean difference in femoral component anteversion between intraoperative and postoperative CT measurements (system accuracy of component anteversion) was 0.8° (SD 1.8°) in the Exeter group and 2.1° (SD 2.3°) in the Accolade II group, respectively (p = 0.020). The mean difference in anteversion between the plan and the postoperative CT measurements (clinical accuracy of femoral component anteversion) was 1.2° (SD 3.6°) in the Exeter group, and 4.2° (SD 3.9°) in the Accolade II group (p = 0.003). No significant differences were found in acetabular component inclination and anteversion; however, the clinical accuracy of combined anteversion was significantly better in the Exeter group (0.6° (SD 3.9°)) than the Accolade II group (3.6° (SD 4.1°)). Conclusion The Mako THA system and software helps surgeons control the femoral component anteversion to achieve the target angle of insertion. The Exeter femoral component, inserted using Mako THA system, showed greater precision for femoral component and combined component anteversion than the Accolade II component. Cite this article: Bone Joint J 2024;106-B(3 Supple A):104–109.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Sugano et al. (2024) studied this question.

synapsesocial.com/papers/68e761d0b6db6435876d7ff1https://doi.org/10.1302/0301-620x.106b3.bjj-2023-0840.r1
Ask AI
Helpful
Bookmark
Share
View Full Paper