October 11, 2025Journal of Refractive Surgery3 citations

Optical Differentiation of Presbyopia-Correcting Intraocular Lenses With Improved Intermediate Vision From a Single Manufacturer

View Full Paper
HSHyeck-Soo SonJohns Hopkins UniversityGŁGrzegorz ŁabuzHeidelberg UniversityZWZhiyi WuApple (Israel)

Key Points

  • All lenses showed comparable optical quality for distance vision, but varied for intermediate vision.
  • The Symfony lens produced a secondary visual acuity peak due to its bifocal design, enhancing some metrics.
  • The Eyhance and PureSee lenses were influenced by pupil size, exhibiting myopic shifts in far focus at smaller sizes.
  • Photic phenomena were minimal with Eyhance and PureSee, while the Symfony lens demonstrated significant halos.

Abstract

Purpose: To compare the visual quality metrics of monofocal and presbyopia-correcting intraocular lenses (IOLs) produced by the same manufacturer (Johnson & Johnson Vision). Methods: This laboratory study analyzed IOLs featuring a TECNIS hydrophobic material, including a monofocal lens (ZCB00), a monofocal lens with improved intermediate vision (Eyhance ICB00), and two different extended-depth-of focus (EDOF) lenses (Symfony and PureSee). An optical bench was used to measure the modulation transfer function (MTF) and United States Air Force (USAF) targets at pupil sizes of 2 to 4.5 mm using polychromatic light and a model cornea with +0.27 µm spherical aberration. Weighted optical transfer function (OTF) was used to simulate the clinical visual acuity and point spread function (PSF) was employed to estimate the light distribution and halo size. Results: All lenses demonstrated comparable optical quality at far distance with simulated visual acuity better than 0.0 logarithm of the minimum angle of resolution (logMAR). However, at the intermediate distance of 67 cm, distinct differences emerged: the Symfony lens showed a secondary visual acuity peak due to its bifocal design, whereas the PureSee lens outperformed Eyhance lens in simulated visual acuity between −1.00 and −2.00 diopters. The Eyhance and PureSee lenses exhibited pupil-dependent behavior, with a myopic shift in far focus at smaller pupil sizes (< 3 mm), whereas the Symfony lens maintained stable performance. Notably, photic phenomena were minimal for the Eyhance and PureSee lenses and similar to the monofocal ZCB00 lens, whereas the Symfony lens displayed pronounced halos. Conclusions: Although all lenses performed well at distance, differences in intermediate vision, pupil dependency, and photic phenomena highlight the importance of individualized IOL selection.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Son et al. (2025) studied this question.

synapsesocial.com/papers/68e9b1d9ba7d64b6fc132fbahttps://doi.org/10.3928/1081597x-20250805-05
Ask AI
Helpful
Bookmark
Share
View Full Paper

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Optical Bench Evaluation of the Latest Refractive Enhanced Depth of Focus Intraocular Lens2024 · 16 citations
  2. 2Clinical outcomes and optical performance of a new segmental refractive extended depth-of-focus intraocular lens2024 · 7 citations
  3. 3Tolerance to refractive error with a new extended depth of focus intraocular lens2024 · 26 citations
  4. 4Symfony Extended Depth of Focus IOL: a Review of Reported Data2017 · 9 citations
  5. 5Visual performance after bilateral implantation of 2 new presbyopia-correcting intraocular lenses: Trifocal versus extended range of vision2017 · 211 citations