Quantification of small (1–10 µm) microplastic particles in soil matrices using automated scanning electron microscopy: possibilities and limitations

View Full Paper
RKRälf KaegiMPMatthias PhilippIJIsabel S. Jüngling

Key Points

  • Automated scanning electron microscopy can detect microplastic particles, achieving around 70% recovery for polyethylene.
  • Monte Carlo simulations indicated optimal SEM conditions with 3 kV of acceleration voltage for better contrast.
  • Using gold coating on polycarbonate membranes improves detection of microplastics in complex soil matrices.
  • Density separation showed negligible presence of PVC in soil suspensions, underlining the complexity of microplastic detection.

Abstract

Abstract Scanning electron microscopy (SEM) provides superb resolution power and allows distinguishing microplastic particles (MPs) from other particles based on elemental ratios derived from energy dispersive x-ray (EDX) analysis. Polycarbonate (PC) membranes are typically used to isolate MPs from suspensions for SEM investigations. This, however, results in poor (atomic weight) contrast between MPs and the underlying PC membrane, challenging automated particle detection. To introduce an elemental contrast between MPs and the PC membrane, a gold (Au) coating was applied to the PC membrane before use. Monte Carlo simulations conducted to optimize operational conditions of the SEM suggest an acceleration voltage of 3 kV in combination with a 40 nm Au coating of PC membranes as most appropriate. Stock suspensions made from polyethylene (PE) fragments, polyvinyl chloride (PVC) fragments, and standard soils, all sieved to 0.45–10 µm, were mixed at different ratios. Selected mixtures were density separated to isolate the MPs from the mixtures. Whereas up to 7% of PE was detected in PVC stocks, likely representing false positives, only negligible amounts of PVC were detected in PE stocks. In soils, neither PE nor PVC was detected in significant amounts. Polyethylene and PVC particles diluted in soil suspensions at particle number ratios of 1:1:2 and 1:1:2000 were density separated, and largely concentration independent recoveries of around 70% and 50% were obtained for PE and PVC, respectively. Good agreement between SEM-EDX and Raman measurements was obtained for MPs, underlining the strength of our approach to address the smallest size fraction of MPs in complex matrices. Graphical abstract

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Kaegi et al. (2025) studied this question.

synapsesocial.com/papers/68f10ecee6a12fd042899897https://doi.org/10.1007/s00216-025-06111-8
Ask AI
Helpful
Bookmark
Share
View Full Paper

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Towards a reference material for microplastics’ number concentration—case study of PET in water using Raman microspectroscopy2024 · 23 citations
  2. 2Computer-controlled scanning electron microscope: Methodologies and application scenarios in atmospheric particle research2023 · 16 citations
  3. 3Microplastic pollution in deep-sea sediments2013 · 1,578 citations
  4. 4Raman microspectroscopy as a tool for microplastic particle analysis2018 · 340 citations
  5. 5Chemical Analysis of Microplastics and Nanoplastics: Challenges, Advanced Methods, and Perspectives2021 · 1,131 citations