Finding an Initial Probe Pose in Teleoperated Robotic Echocardiography via 2D LiDAR-Based 3D Reconstruction

View Full Paper
MRMariadas Capsran RoshanEHEdgar M. HidalgoMIMats Isaksson

Key Points

  • The proposed 2D LiDAR system significantly reduces the time to find an initial probe pose in robotic echocardiography.
  • With a mean surface error of 2.78 mm, the approach demonstrates high accuracy compared to traditional methods.
  • The method integrates plane-based extrinsic calibration, achieving a root mean square residual of 1.8 mm.
  • Human trials reveal that the probe points are consistently within 20-30 mm of the clinically defined starting location.

Abstract

Echocardiography is a key imaging modality for cardiac assessment but remains highly operator-dependent, and access to trained sonographers is limited in underserved settings. Teleoperated robotic echocardiography has been proposed as a solution; however, clinical studies report longer examination times than manual procedures, increasing diagnostic delays and operator workload. Automating non-expert tasks, such as automatically moving the probe to an ideal starting pose, offers a pathway to reduce this burden. Prior vision- and depth-based approaches to estimate an initial probe pose are sensitive to lighting, texture, and anatomical variability. We propose a robot-mounted 2D LiDAR-based approach that reconstructs the chest surface in 3D and estimates the initial probe pose automatically. To the best of our knowledge, this is the first demonstration of robot-mounted 2D LiDAR used for 3D reconstruction of a human body surface. Through plane-based extrinsic calibration, the transformation between the LiDAR and robot base frames was estimated with an overall root mean square (RMS) residual of 1.8 mm and rotational uncertainty below 0.2°. The chest front surface, reconstructed from two linear LiDAR sweeps, was aligned with non-rigid templates to identify an initial probe pose. A mannequin-based study assessing reconstruction accuracy showed mean surface errors of 2.78 +/- 0.21 mm. Human trials (N=5) evaluating the proposed approach found probe initial points typically 20-30 mm from the clinically defined initial point, while the variation across repeated trials on the same subject was less than 4 mm.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Roshan et al. (2025) studied this question.

synapsesocial.com/papers/68f5fcce8d54a28a75cf1d65https://doi.org/10.48550/arxiv.2509.24867
Ask AI
Helpful
Bookmark
Share
View Full Paper

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Teleoperated Probe Manipulator for Prone-Position Echocardiography Examination2022 · 5 citations
  2. 2Initial Transducer Placement for Automated Robotic Ultrasound Examinations2025 · 1 citations
  3. 3A robot-assisted lung ultrasound system for standardized scanning2025
  4. 4Iterative trocar-constrained laparoscope viewpoint search for enhanced surface coverage in 3D reconstruction2026
  5. 5Robotic assisted echocardiography, a pilot clinical validation study2026