PulseExploreJournal ClubDebatesTrendingResearchersJournals
Instagram
HomeExploreJournal ClubTrending
Synapse
⌘+K
Synapse
December 5, 2025Frontiers in Transplantation7 citationsOpen Access

Comparative evaluation of five tacrolimus assays in transplant recipients: implications for optimizing therapeutic drug monitoring

View Full Paper
JAJ. Alex

Key Points

  • Strong correlations among tacrolimus assays reveal significant variability in monitoring methods.
  • Immunoassays showed moderate agreement with LC-MS/MS methods, indicating potential clinical implications.
  • Deming regression and Bland–Altman analysis underscore differences in tacrolimus assay performance.
  • Findings support the need for standardized monitoring practices to ensure effective immunosuppressive therapy.

Abstract

Tacrolimus is a widely used immunosuppressive therapy in transplant recipients, but its narrow therapeutic index necessitates accurate monitoring. Tacrolimus levels can be quantified using immunoassays (IAs) and liquid chromatography-tandem mass spectrometry (LC-MS/MS), however, differences between these methods may influence clinical decision-making. In this study, we compared two IAs, i.e., chemiluminescent (CMIA) and electrochemiluminescent (ECLIA), with three LC-MS/MS assays in 181 clinical specimens. When compared with the overall mean concentration, all five assays showed strong correlations, though with variability across methods: three LC-MS/MS assays demonstrated correlation coefficients of 0.9927, 0.9612, and 0.9920, while two immunoassays yielded coefficients of 0.9938 and 0.9857. Deming regression analysis revealed slopes of 0.96, 0.94, and 0.93 for the three LC-MS/MS, while the immunoassays showed higher slopes of 1.032 (ECLIA) and 1.21 (CMIA). Bland–Altman analysis indicated systematic underestimation by the LC-MS/MS methods (–7.5%, −18.7%, and −8%) and overestimation by the immunoassays (ECLIA +9.7%, CMIA +18.4%), relative to the overall mean. The two immunoassays showed only moderate agreement with each other (slope = 0.85, intercept = 0.49), and even the LC-MS/MS assays were not fully concordant. Among 47 patients within 3 months post-transplantation and 134 patients beyond 3 months, clinically relevant discrepancies (≥2 ng/ml) between LC-MS/MS and immunoassay results were observed in 13 patients (28%) and 49 patients (37%), respectively. These findings underscore the substantial impact of assay-dependent variability on tacrolimus monitoring and emphasize the need for standardized laboratory practices as well as assay-specific therapeutic ranges to prevent underexposure with rejection or overexposure with toxicity.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

J. Alex (2025) studied this question.

synapsesocial.com/papers/693231118e51979591dcdfd8https://doi.org/10.3389/frtra.2025.1716789
Ask AI
Helpful
Bookmark
Share
View Full Paper

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Impact of IL-6 and IL-10 genotypes on tacrolimus dose requirements in kidney transplant recipients: Monte Carlo analysis2024 · 2 citations
  2. 2Combined Effect of Inter- and Intrapatient Variability in Tacrolimus Exposure on Graft Impairment Within a 3-Year Period Following Kidney Transplantation: A Single-Center Experience2020 · 12 citations
  3. 3Comparison of LC-MS/MS and chemiluminescent immunoassays for immunosuppressive drugs reveals organ dependent variation in blood cyclosporine a concentrations2020 · 19 citations
  4. 4The calcineurin inhibitor tacrolimus activates the renal sodium chloride cotransporter to cause hypertension2011 · 385 citations
  5. 5Calcineurin Phosphatase Activity and Immunosuppression. A Review on the Role of Calcineurin Phosphatase Activity and the Immunosuppressive Effect of Cyclosporin A and Tacrolimus2003 · 96 citations