January 10, 2026Hypertension Research0 citationsOpen Access

Comparison of two validated oscillometric devices in a home-like setup reveals pronounced blood pressure differences and reduced precision

JZJoachim ZahndBTBrian ThompsonPRPierre-Antonin Rigon

Key Result

The wrist monitor overdiagnosed hypertension, identifying nearly twice as many hypertensive individuals compared to the upper-arm device.

Key Points

  • This research aims to assess the accuracy and precision of two oscillometric devices for measuring blood pressure in a home-like environment.
  • Compared wrist-based and upper-arm blood pressure monitors in a home-like setup
  • 121 participants performed four concurrent blood pressure measurements
  • Conducted a crossover of devices between arms to compare results
  • Evaluated bias and precision of each device's measurements
  • Wrist monitor readings were consistently higher than upper-arm measurements
  • Both devices showed increased bias at higher blood pressure levels
  • Moderate agreement in hypertension classification (Cohen’s kappa=0.58)
  • Wrist monitor over-diagnosed hypertension and had greater variability compared to upper-arm device

Study Design

Type

Observational (n=121)

Multicenter

No

Structured PICO

Does a wrist-worn oscillometric blood pressure monitor provide accurate blood pressure measurements and hypertension classification compared to an upper-arm monitor when placed by patients in a home-like setup?

P
Population
121 adult participants across wide blood pressure ranges, mean age 48.5 years, 51.5% female, evaluated in a dedicated outpatient consultation room mimicking a home-like setup.
I
Intervention
Wrist-worn oscillometric blood pressure monitor placed by participants themselves.
C
Comparator
Reference upper-arm oscillometric blood pressure monitor placed by participants themselves.
O
Outcome
Agreement, bias, and precision of systolic and diastolic blood pressure measurements, and hypertension classification (threshold 135/85 mmHg).surrogate

Wrist-worn blood pressure monitors exhibit significant bias, reduced precision, and tend to overdiagnose hypertension compared to upper-arm monitors when used by patients in home-like settings.

Main Result

Effect estimate: null (95% CI null)

Absolute Event Rate: 47.1% vs 28.1%

p-value: p=<0.0001

Limitations

  • Participants were responsible for device placement, which may introduce bias.
  • Absence of office BP measurements to compare results in a controlled setting.
  • Absence of office blood pressure measurement that would have added insight of the setting effect
  • Absence of sensor may have introduced a bias through incorrect positioning of the wrist device
  • Did not record the error of position during the study

Abstract

Abstract Home blood pressure monitoring (HBPM) is essential for long-term hypertension management, but its accuracy and reliability is questionable due to user inconsistencies and non-standard usage conditions. This study compared two validated HBPM devices – a wrist-based and a reference upper-arm monitor – in a home-like setup where participants placed the devices themselves. A total of 121 participants underwent four concurrent blood pressure (BP) measurements, with two taken in each arm, followed by a crossover of devices between arms after the initial two readings. Wrist-derived BP readings were higher than upper-arm measurements. With higher blood pressure levels, both devices exhibited greater bias, accompanied by reduced precision in systolic BP measurements. Additionally, hypertension classification showed only moderate agreement (Cohen’s kappa=0.58). The wrist monitor tended to over-diagnose hypertension and exhibited greater variability than the upper-arm device. These findings highlight the need for more guidance and education as well as critical evaluation of home blood pressure measurements provided by patients.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Zahnd et al. (2026) conducted an observational in Hypertension (n=121). Omron BP RS4 (wrist monitor) vs. Microlife WatchBP Home (upper-arm monitor) was evaluated on Systolic blood pressure difference between wrist and upper-arm monitors (null, 95% CI null, p=<0.0001). The wrist monitor overdiagnosed hypertension, identifying nearly twice as many hypertensive individuals compared to the upper-arm device.

synapsesocial.com/papers/696321b391e05aa366cb7ed1https://doi.org/10.1038/s41440-025-02514-3
Ask AI
Helpful
Bookmark
Share
View Full Paper

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Poor Reliability of Wrist Blood Pressure Self-Measurement at Home2016 · 59 citations
  2. 2Comparison of Home and Office Blood Pressure Devices in the Clinical Setting2023 · 8 citations
  3. 3Impact of Miscuffing During Home Blood Pressure Measurement on the Prevalence of Masked Hypertension2013 · 21 citations
  4. 4Validation of two automatic devices, Omron HEM-6232T and HEM-6181, for self-measurement of blood pressure at the wrist according to the ANSI/AAMI/ISO 81060-2:2013 protocol and the European Society of Hypertension International Protocol revision 20102019 · 11 citations
  5. 5Validation of the Microlife WatchBP Home device for self home blood pressure measurement according to the International Protocol2007 · 98 citations