January 10, 2026Pacing and Clinical Electrophysiology0 citations

Horizontal Axis Affects Paced QRS Duration More Than Vertical Axis at Leadless Pacemaker Implantation Site

View Full Paper
HSHiroko SatoYHYuichi HoriYKYuta Kimura

Key Result

The horizontal position in the RV septum correlates more strongly with paced QRS duration (R = 0.59) compared to the vertical position (R = 0.03).

Key Points

  • This research aims to determine how the horizontal and vertical axes affect paced QRS duration during leadless pacemaker implantation.
  • Fifty-four patients underwent leadless pacemaker implantation guided by three-dimensional rotational angiography.
  • Implantation sites were assessed using horizontal and vertical axes within the right ventricle (RV).
  • Paced QRS durations were measured intraoperatively using electrocardiograms.
  • Echocardiography was performed 12 months later to evaluate progression of cardiomyopathy.
  • All 54 leadless pacemakers were successfully implanted in the septum.
  • Mean horizontal and vertical implantation positions were measured as 69.7% and 66.5%, respectively.
  • Mean paced QRS duration was recorded at 148 ms.
  • Strong correlation found between horizontal position and paced QRS duration (R = 0.59); vertical position showed no correlation (R = 0.03).
  • No patients experienced progression of cardiomyopathy during follow-up.

Structured PICO

Does the specific implantation site (horizontal vs vertical axis) of a leadless pacemaker in the right ventricular septum affect paced QRS duration in patients undergoing implantation?

P
Population
54 patients undergoing leadless pacemaker implantation.
I
Intervention
Leadless pacemaker implantation in the right ventricular septum using three-dimensional rotational angiography (3D-RA) guidance.
O
Outcome
Paced QRS duration (p-QRS) measured intraoperatively.surrogate

The horizontal position of a leadless pacemaker in the right ventricular septum is a stronger determinant of paced QRS duration than the vertical position.

Abstract

ABSTRACT Background Leadless pacemaker (LP) implantation at the apical septum has been linked to pacing‐induced cardiomyopathy (PICM). This study aimed to examine the precise position for LP implantation in the RV septum using the three‐dimensional rotational angiography (3D‐RA) system. The accurate implantation sites were visualized through the fluoroscopic image, and the relationship to the measured QRS duration during pacing (p‐QRS duration) was investigated to evaluate their potential association with cardiomyopathy. Methods Fifty‐four patients underwent LP implantation with 3D‐RA guidance. Implantation sites were assessed using the horizontal axis from the apex to the center of the tricuspid valve and the vertical axis from the inferior border of the apex to the outflow tract transition. The p‐QRS durations were measured intraoperatively using electrocardiograms. Echocardiography was performed 12 months later to assess PCIM progression. Results All 54 LPs were successfully implanted in the septum. Mean horizontal and vertical positions within the RV were 69.7 ± 10.5% and 66.5 ± 13.1%, respectively. Mean p‐QRS duration was 148 ± 11 ms. The horizontal position was correlated with QRS duration ( R = 0.59), while the vertical position showed no correlation ( R = 0.03). The native QRS (n‐QRS) duration was also an independent predictor of prolonged p‐QRS duration ( p = 0.018). No patients developed cardiomyopathy progression on follow‐up, and no association with p‐QRS duration was observed. Conclusion In LP implantation, the horizontal position within the RV septum correlates more strongly with p‐QRS duration than vertical position. Prolonged n‐QRS duration also prolonged p‐QRS duration; however, further research is needed to evaluate the relationship between the implanted position and the occurrence of PICM.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Sato et al. (2025) studied this question. The horizontal position in the RV septum correlates more strongly with paced QRS duration (R = 0.59) compared to the vertical position (R = 0.03).

synapsesocial.com/papers/6963223291e05aa366cb8c6dhttps://doi.org/10.1111/pace.70107