January 22, 2026Advanced Functional Materials0 citations

An Implantable Flexo‐Ferroelectric Sensor Enabled by a Bioresorbable Nanocrystal‐Nanofiber Heterostructure for Monitoring Internal Soft Tissue Strains

View Full Paper
SYShuang YuWZWenjun ZhaoGCGuang Chen

Key Points

  • This research aims to develop a flexible, sensitive, and biodegradable sensor for monitoring internal soft tissue strains.
  • Engineered a flexo-ferroelectric nanocomposite using β-glycine nanocrystals in PLLA nanofibers.
  • Utilized an electric field-assisted nanoconfinement strategy for in situ growth and stabilization.
  • Conducted in vivo validation in rodent and rabbit models for continuous cardiovascular monitoring.
  • Achieved a piezoelectric coefficient of 246 pC/N and a sensitivity of 205 mV/kPa.
  • Captured strain across soft tissues with a detection range of 3.5-110 kPa.
  • Successfully distinguished between metabolic and mechanical modes of acute heart failure.

Abstract

ABSTRACT Implantable sensors are crucial for real‐time monitoring of internal soft tissue biomechanics, yet existing systems rarely combine high sensitivity, flexibility, and biodegradability. Here, we present an implantable sensor by engineering a flexo‐ferroelectric nanocomposite, where ferroelectric β‐glycine nanocrystals are in situ‐grown and stabilized along poly( L ‐lactic acid) (PLLA) nanofibers via an electric field‐assisted nanoconfinement strategy. This approach first uncovers flexoelectricity in PLLA nanofibers and template‐freely stabilizes pure β‐glycine nanocrystals without brittle polymer matrices. These components form a structurally coupled, self‐assembled interface, creating a co‐active charge generation network that integrates flexoelectric and ferroelectric effects new mechanism not previously demonstrated. Our sensor achieves unprecedented electromechanical performance, including piezoelectric coefficient of 246 pC/N, sensitivity of 205 mV/kPa, detection range of 3.5‐110 kPa, and tissue‐matched compliance. The integrated performance surpasses all reported bioresorbable sensors and commercial polyvinylidene fluoride‐based devices. Functionally, it demonstrates layer‐specific compatibility across soft tissues, capturing strain from surface to deep regions with minimal signal disruption. In vivo validation in rodent and rabbit models enables continuous cardiovascular monitoring and, notably, discriminates between metabolic and mechanical modes of acute heart failure. This work establishes a universal sensing platform for high‐fidelity soft tissue strain mapping, with broad implications for personalized mechano‐diagnostics across multiscale physiological systems.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Yu et al. (2026) studied this question.

synapsesocial.com/papers/6971be8d642b1836717e328ehttps://doi.org/10.1002/adfm.202529162
Ask AI
Helpful
Bookmark
Share
View Full Paper

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Ag–thiolate interactions to enable an ultrasensitive and stretchable MXene strain sensor with high temporospatial resolution2024 · 93 citations
  2. 2A Biodegradable Piezoelectric Sensor for Real‐Time Evaluation of the Motor Function Recovery After Nerve Injury2024 · 82 citations
  3. 3A dual‐range Janus‐structure pressure sensor with broad detection range and high resolution combining triboelectricity and piezoelectricity2024 · 25 citations
  4. 4Iatrogenic Injury From Vascular Access and Endovascular Procedures2011 · 49 citations
  5. 5Hyper strength, high sensitivity integrated wearable signal sensor based on non-covalent interaction of an ionic liquid and bacterial cellulose for human behavior monitoring2024 · 38 citations